Usefulness of Sealants for Dural Closure: Evaluation in an In Vitro Model

Publication date

2018-10

Authors

van Doormaal, Tristan
Kinaci, Ahmet
van Thoor, Sander
Redegeld, Saskia
Bergmann, WilhelminaISNI 0000000419550328
Van Der Zwan, Albert

Editors

Advisors

Supervisors

Document Type

Article
Open Access logo

License

taverne

Abstract

BACKGROUND: Cerebrospinal fluid (CSF) leakage occurs in 4% to 32% of cranial surgeries and is associated with significant patient burden and expense. The use of sealant as an adjunct to primary dural closure is assumed to help prevent CSF leakage. OBJECTIVE: To examine the utility of different sealants for dural closure using an in Vitro model. METHODS: We evaluated 9 commonly used dural sealants, including Tachosil (Takeda Inc, Osaka, Japan), Adherus (Hyperbranch Inc, Durham, North Carolina), Duraform (Codman, Raynham, Massachusetts), Tissudura (Baxter, Deerfield, Illinois), Hemopatch (Baxter), TissuePatchDural (Tissuemed, Leeds, United Kingdom), Tisseel (Baxter), Duragen Secure (Integra, Plainsboro, New Jersey), and Duraseal, (Integra). Sealants were tested in 2 novel in Vitro setups using fresh porcine dura: the first tested the acute burst pressure of a sealed 3-mm gap, while the second examined resistance to a pressure wave mimicking intracranial pressure for 72 h. RESULTS: Adherus showed the highest mean burst pressure (87 ± 47 mmHg) followed by Tachosil (71 ± 16 mmHg) and Duraseal (51 ± 42 mmHg); these were the only 3 sealants showing burst pressures above normal physiological intracranial pressure. In the 72-h setup, only Adherus and Duraseal maintained appropriate sealing for the duration of the experiment. Tachosil released from the dura after 1.4 h (95% confidence interval, -1.8-4.7). CONCLUSION: Given the high cost of sealants and the results of this study, we advocate a critical attitude toward sealant application as an adjunct to classic dural closure.

Keywords

Cerebrospinal fluid leakage, Dura mater, Dural closure, Taverne

Citation

van Doormaal, T, Kinaci, A, van Thoor, S, Redegeld, S, Bergmann, W & Van Der Zwan, A 2018, 'Usefulness of Sealants for Dural Closure : Evaluation in an In Vitro Model', Operative neurosurgery (Hagerstown, Md.), vol. 15, no. 4, pp. 425–432. https://doi.org/10.1093/ons/opx260