Arterial stiffness as underlying mechanism of disagreement between an oscillometric blood pressure monitor and a sphygmomanometer

Publication date

2000-01-01

Authors

Van Popele, Nicole M.
Bos, Willem Jan W.
De Beer, Nicole A.M.
Van Der Kuip, Deirdre A.M.
Hofman, A.
Grobbee, RickORCID 0000-0003-4472-4468ISNI 0000000030206553
Witteman, Jacqueline C.M.

Editors

Advisors

Supervisors

Document Type

Article

Collections

License

Abstract

Oscillometric blood pressure devices tend to overestimate systolic blood pressure and underestimate diastolic blood pressure compared with sphygmomanometers. Recent studies indicate that discrepancies in performance between these devices may differ between healthy and diabetic subjects. Arterial stiffness in diabetics could be the underlying factor explaining these differences. We studied differences between a Dinamap oscillometric blood pressure monitor and a random-zero sphygmomanometer in relation to arterial stiffness in 1808 healthy elderly subjects. The study was conducted within the Rotterdam Study, a population-based cohort study of subjects aged 55 years and older. Systolic and diastolic blood pressure differences between a Dinamap and a random-zero sphygmomanometer were related to arterial stiffness, as measured by carotid-femoral pulse wave velocity. Increased arterial stiffness was associated with higher systolic and diastolic blood pressure readings by the Dinamap compared with the random-zero sphygmomanometer, independent of age, gender, and average mean blood pressure level of both devices. The β-coefficient (95% CI) was 0.25 (0.00 to 0.50) mm Hg/(m/s) for the systolic blood pressure difference and 0.35 (0.20 to 0.50) mm Hg/(m/s) for the diastolic blood pressure difference. The results indicate that a Dinamap oscillometric blood pressure device, in comparison to a random-zero sphygmomanometer, overestimates systolic and diastolic blood pressure readings in subjects with stiff arteries.

Keywords

Arterial stiffness, Blood pressure monitoring, Oscillometry, Sphygmomanometry, Internal Medicine

Citation

Van Popele, N M, Bos, W J W, De Beer, N A M, Van Der Kuip, D A M, Hofman, A, Grobbee, D E & Witteman, J C M 2000, 'Arterial stiffness as underlying mechanism of disagreement between an oscillometric blood pressure monitor and a sphygmomanometer', Hypertension, vol. 36, no. 4, pp. 484-488. https://doi.org/10.1161/01.HYP.36.4.484