Monitoring of absolute mirror alignment at COMPASS RICH-1 detector

Files

Access status: Embargo until 2050-01-01 , 4_main.pdf (292.16 KB)

Publication date

2014-12-01

Authors

Alexeev, M.
Birsa, R.
Bradamante, F.
Bressan, A.
Chiosso, M.
Ciliberti, P.
Dalla Torre, S.
Denisov, O.
Duic, V.
Ferrero, A.

Editors

Advisors

Supervisors

Document Type

Article

License

Abstract

The gaseous COMPASS RICH-1 detector uses two spherical mirror surfaces, segmented into 116 individual mirrors, to focus the Cherenkov photons onto the detector plane. Any mirror misalignment directly affects the detector resolution. The on-line Continuous Line Alignment and Monitoring (CLAM) photogrammetry-based method has been implemented to measure the alignment of individual mirrors which can be characterized by the center of curvature. The mirror wall reflects a regular grid of retroreflective strips placed inside the detector vessel. Then, the position of each mirror is determined from the image of the grid reflection. The images are collected by four cameras. Any small mirror misalignment results in changes of the grid lines' positions in the image. The accuracy limits of the CLAM method were checked by laser interferometry and are below 0.1 mrad.

Keywords

Cherenkov radiation, Mirror alignment, Photogrammetry, RICH, Instrumentation, Nuclear and High Energy Physics

Citation

Alexeev, M, Birsa, R, Bradamante, F, Bressan, A, Chiosso, M, Ciliberti, P, Dalla Torre, S, Denisov, O, Duic, V, Ferrero, A, Finger, M, Finger, M, Gayde, J C, Giorgi, M, Gobbo, B, Levorato, S, Maggiora, A, Martin, A, Menon, G, Panzieri, D, Pesaro, G, Polak, J, Rocco, E, Sbrizzai, G, Schiavon, P, Slunecka, M, Sozzi, F, Steiger, L, Sulc, M, Takekawa, S & Tessarotto, F 2014, 'Monitoring of absolute mirror alignment at COMPASS RICH-1 detector', Nuclear Instruments and Methods in Physics Research. Section A, vol. 766, pp. 208-211. https://doi.org/10.1016/j.nima.2014.06.066