Liquid chromatography-tandem mass spectrometric assay for the T790M mutant EGFR inhibitor osimertinib (AZD9291) in human plasma
Files
Publication date
2016-09-15
Editors
Advisors
Supervisors
Document Type
Article
Metadata
Show full item recordCollections
License
taverne
Abstract
A method for the quantitative analysis by ultra-performance liquid chromatography-tandem mass spectrometry of the highly selective irreversible covalent inhibitor of EGFR-TK, osimertinib in human plasma was developed and validated, using pazopanib as an internal standard. The validation was performed in a range from 1 to 1000ng/ml, with the lowest level corresponding to the lower limit of quantitation. Gradient elution was performed on a 1.8μm particle trifunctional bonded C18 column by 1% (v/v) formic acid in water, and acetonitrile as mobile phase. The analyte was detected in the selected reaction monitoring mode of a triple quadrupole mass spectrometer after positive ionization with the heated electrospray interface. Within-day precisions ranged from 3.4 to 10.3%, and between-day precisions from 3.8 to 10.4%, accuracies were 95.5-102.8%. Plasma (either lithium heparin or sodium EDTA) pretreatment was performed by salting-out assisted liquid-liquid extraction using acetonitrile and magnesium sulfate. This method was used to analyze the osimertinib blood plasma levels of five adult patients with metastatic T790M mutated non-small cellular lung carcinoma for therapeutic drug monitoring purposes.
Keywords
Osimertinib, Non-small cell lung carcinoma, LC–MS/MS, Human plasma, Salting-out assisted liquid–liquid extraction, Taverne
Citation
Rood, J J M, van Bussel, M T J, Schellens, J H M, Beijnen, J H & Sparidans, R W 2016, 'Liquid chromatography-tandem mass spectrometric assay for the T790M mutant EGFR inhibitor osimertinib (AZD9291) in human plasma', Journal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences, vol. 1031, pp. 80-85. https://doi.org/10.1016/j.jchromb.2016.07.037